Instrument setup
R&S
®
MXO 5 Series
94User Manual 1802.3369.02 ─ 02
1. Warm up the instrument before you start the self-alignment. The minimum warm-up
time is indicated in the data sheet.
2. Remove the probes from the input connectors.
3. Open "Menu" > "Settings" > "Maintenance".
4. In the "Alignment" tab, tap "Start alignment".
The alignment is performed, the process can take several minutes. A message box
informs you about the running process, wait until this message box closes.
The overall pass/fail result is shown in the "Overall alignment state" field.
5.7.3 Power management
Access: "Menu" > "Settings" > "Maintenance" > "Power mgmt."
Auto power up
If enabled, the instrument powers up automatically when it is connected to the mains
voltage.
Remote command:
SYSTem:APUP on page 460
5.7.4 Service
Access: "Menu" > "Settings" > "Maintenance" > "Service"
Maintenance settings