Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 461

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
461User Manual 1802.3369.02 ─ 02
Manual operation: See "Run Single" on page 107
STOP
Stops the running acquisition.
Usage: Event
Asynchronous command
Manual operation: See "Run/Stop" on page 107
17.8.2 Horizontal setup
AUToscale.....................................................................................................................461
TIMebase:SCALe...........................................................................................................461
TIMebase:RANGe..........................................................................................................461
TIMebase:DIVisions?......................................................................................................462
TIMebase:HORizontal:POSition.......................................................................................462
TIMebase:REFerence.....................................................................................................462
AUToscale
Performs an autoset process: analyzes the enabled channel signals, and obtains
appropriate horizontal, vertical, and trigger settings to display stable waveforms.
Rohde & Schwarz does not recommend using the autoset in remote control. To adjust
the oscilloscope remotely, especially for automated testing applications, use the
remote commands that adjust the horizontal, vertical and trigger settings.
Usage:
Event
Asynchronous command
TIMebase:SCALe <TimebaseScale>
Sets the horizontal scale, the time per division, for all waveforms in the time domain,
for example, channel and math waveforms.
Parameters:
<TimebaseScale> Range: 200E-12 to 10E3
Increment: 1E-12
*RST: 20E-9
Default unit: s/div
Usage: Asynchronous command
Manual operation: See "Time scale" on page 101
TIMebase:RANGe <TimebaseRange>
Sets the time of one acquisition, which is the time across the 10 divisions of the dia-
gram: Acquisition time = Time scale * 10 divisions.