Trigger
R&S
®
MXO 5 Series
139User Manual 1802.3369.02 ─ 02
If an external trigger signal is connected to the Trigger In connector, and the trigger
source is set to "Extern", the analog edge trigger is available. Triggering on an external
source is only possible if you trigger on a single event but not for sequences.
The "Slope" and "Level" are the same settings as for the edge trigger, see:
"Level" on page 137
"Slope" on page 138
Specific settings for the analog edge trigger are the following:
Coupling
Sets the connection of the external trigger signal, i.e. the input impedance and a termi-
nation. The coupling determines what part of the signal is used for triggering.
"DC 50
"
Connection with 50 Ω termination, passes both DC and AC compo-
nents of the signal.
Trigger types