Remote control commands
R&S
®
MXO 5 Series
683User Manual 1802.3369.02 ─ 02
Return values:
<FrameByteStart> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:I2C:FRAMe<fr>:BYTE<o>:VALue?
Returns the data value of the specified byte.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<o> index of the byte number
Return values:
<Value> Range: 0 to 255
Increment: 1
*RST: 0
Usage: Query only
Asynchronous command
SBUS<sb>:I2C:FRAMe<fr>:DATA?
Returns the data words of the specified frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Example:
BUS:I2C:FRAMe4:DATA?
<-- 3,74,164,18
Usage: Query only
Asynchronous command
SBUS<sb>:I2C:FRAMe<fr>:RWBStart?
Returns the start time of the R/W bit
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame