262User Manual 1802.3369.02 ─ 02
11.2.4.1 Deskew settings
Skew channel
Sets user-defined skew settings for the channel connected to the current probe.
Alternatively, you can start an "Auto Deskew".
Auto Deskew
Starts an auto deskew. It is used to determine the trigger and scaling properties.
For an auto deskew, only Rohde & Schwarz probes are supported.
Make sure that the probes are configured correctly before you start the deskewing.
After the deskew, the result values are written in the "Skew" value of the corresponding
channel.
If there are more than two active channels, you can perform automatic deskew step by
step and exchange the probes connected fixture between the steps.
Remote command:
POWer<m>:ASKew[:EXECute] on page 620
11.2.5 Offset compensation
The offset compensation adjustments are applied to all power analysis measurements.
Power analysis (option R&S
MXO5-K31)