Network operation and remote control
R&S
®
MXO 5 Series
411User Manual 1802.3369.02 ─ 02
Changing the LAN parameters requires the "Web Ifc Password" password, which is set
on the instrument.
IP configuration
The "LAN Configuration > IP configuration" webpage displays all mandatory LAN
parameters and allows their modification. Changing the LAN parameters requires the
"Web Ifc Password" password, which is set on the instrument.
The "IP address mode" controls how the IP address for the instrument gets assigned.
For the manual configuration mode, at least the static IP address, subnet mask, and
default gateway are used to configure the LAN. The automatic configuration mode
uses DHCP server or Dynamic Link Local Addressing to obtain the instrument IP
address.
The same settings are available on the instrument under "Menu" > "Settings" > "Sys-
tem" > "Network" > "Advanced".
Advanced configuration
The "LAN Configuration > Advanced configuration" parameters are used as follows:
mDNS and DNS-SD are two additional protocols: Multicast DNS and DNS Service
Discovery. They are used for device communication in zero configuration networks
working without DNS and DHCP.
"ICMP ping" must be enabled to use the ping utility.
"VXI-11" is the protocol that is used to detect the instrument in the LAN.