Remote control commands
R&S
®
MXO 5 Series
516User Manual 1802.3369.02 ─ 02
Parameters:
<Mode> OFF | LFReject | RFReject
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Trigger filter mode" on page 156
TRIGger:LFReject <Bandwidth>
Sets the limit frequency limit for the highpass filter of the trigger signal. Frequencies
lower than this value are rejected, higher frequencies pass the filter.
Parameters:
<Bandwidth> Range: 50 kHz
*RST: 50 kHz
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "LF reject BW" on page 157
TRIGger:RFReject <Bandwidth>
Sets the limit frequency limit for the lowpass filter of the trigger signal. Frequencies
higher than this value are rejected, lower frequencies pass the filter.
Parameters:
<Bandwidth> Range: 1E+3 to 500E+6
Increment: 1000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "RF reject BW" on page 157
17.9.13 Actions on trigger
TRIGger:ACTions:OUT:STATe......................................................................................... 516
TRIGger:ACTions:OUT:SOURce......................................................................................517
TRIGger:ACTions:OUT:DELay.........................................................................................517
TRIGger:ACTions:OUT:PLENgth..................................................................................... 517
TRIGger:ACTions:OUT:POLarity......................................................................................517
TRIGger:ACTions:OUT:STATe <State>
Activates the outgoing pulse on the [Trigger Out] connector on the rear panel.
Parameters:
<State> ON | OFF
*RST: OFF