Waveform analysis
R&S
®
MXO 5 Series
180User Manual 1802.3369.02 ─ 02
Enhancement mode
The enhancement mode is taken from the source waveform of the reference. The
enhancement mode is set automatically by the instrument and depends on the sample
rate. As long as the waveform sample rate is not higher than the ADC sample rate, the
instrument works automatically in real time mode (RT). At higher waveform sample
rates, the instrument changes to interpolated time mode (IT), where interpolation adds
waveform points between the captured ADC samples by various mathematical meth-
ods, see Interpolation.
Waveform arithmetic
The waveform arithmetic is taken from the source waveform of the reference, and it
cannot be changed for the reference. Waveform arithmetic builds the resulting wave-
form from several consecutive acquisitions of the signal. The arithmetic methods are
envelope and average, which are set in the Acquisition mode.
Restore settings
Applies the original settings of the reference waveform to the horizontal and vertical
settings of the selected waveform.
Remote command:
REFCurve<rc>:RESTore on page 535
8.4.2.3 Reference waveform vertical
Access: [Ref] key > "Vertical" tab
Mode
Selects the type of vertical settings:
"Coupled to
source"
Vertical position and scale of the source are used.
"Independent"
Scaling and position can be set specific to the reference waveform.
Remote command:
REFCurve<rc>:VMODe on page 541
Reference waveforms