Remote control commands
R&S
®
MXO 5 Series
482User Manual 1802.3369.02 ─ 02
NOACtion
Nothing is started on pressing the micro button.
FTRiglevel
Sets the trigger level automatically to 0.5 * (MaxPeak – Min-
Peak). The function is not available for an external trigger
source.
PRSetup
Opens the "Probes Setup" dialog box.
*RST: RCONtinuous
Usage: Asynchronous command
Manual operation: See "MicroButton" on page 124
PROBe<ch>:SETup:ACCoupling <ProbeCouplingAC>
Enables AC coupling in R&S RTZPR power rail probes, which removes DC and very
low-frequency components. The R&S RTZPR probe requires 50 Ω input termination,
for which the channel AC coupling is not available. The probe setting allows AC cou-
pling also at 50 Ω inputs.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<ProbeCouplingAC> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "AC Coupling" on page 127
PROBe<ch>:SETup:CMOFfset <CMOffset>
Sets the common-mode offset to compensate for a common DC voltage that is applied
to both input sockets (referenced to the ground socket). The setting is available for
Rohde & Schwarz differential probes.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<CMOffset> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "CM offset" on page 126