Loading ...
Loading ...
Loading ...
Trigger
R&S
®
MXO 5 Series
138User Manual 1802.3369.02 ─ 02
● Interval trigger....................................................................................................... 148
● Slew rate trigger....................................................................................................150
● Line trigger............................................................................................................ 152
7.3.1 Edge trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Edge"
The edge trigger is the most common trigger type. The trigger occurs when the signal
from the trigger source passes the trigger level in the specified direction (slope).
Level
See "Level" on page 137.
Remote command:
TRIGger:EVENt<ev>:LEVel<n>[:VALue] on page 493
Slope
Sets the edge direction for the trigger.
"Positive"
Selects the rising edge, which is a positive voltage change.
"Negative"
Selects the falling edge, which is a negative voltage change.
"Either"
Selects the rising and falling edge. After starting an acquisition, the
instrument triggers on the first identified edge.
Remote command:
TRIGger:EVENt<ev>:EDGE:SLOPe on page 494
7.3.2 Edge trigger on external trigger source
Access: "Menu" > "Trigger" > "Setup" tab > "Source" = "Extern" > "Type = Analog
Edge"
Trigger types
Loading ...
Loading ...
Loading ...