Remote control commands
R&S
®
MXO 5 Series
594User Manual 1802.3369.02 ─ 02
CALCulate:SPECtrum<sp>:STATe <State>
Enables the spectrum.
Suffix:
<sp>
.
1...4, index of the spectrum
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Display" on page 222
CALCulate:SPECtrum<sp>:THReshold <Threshold>
Sets an absolute threshold as an additional condition for the peak search. Only peaks
that exceed the threshold are detected.
Suffix:
<sp>
.
1...4, index of the spectrum
Parameters:
<Threshold> Range: -500 to 500
Increment: 1
*RST: -70
Default unit: dBm
Usage: Asynchronous command
Manual operation: See "Threshold" on page 192
CALCulate:SPECtrum<sp>:PEXCursion <PeakExcursion>
Defines a relative threshold, the minimum level value by which the waveform must rise
by to be considered as a peak. For the peak to be considered, the excursion threshold
must be met both on the left and the right sides of the peak.
To avoid identifying noise peaks, enter a peak excursion value that is higher than the
noise levels.
Suffix:
<sp>
.
1...4, index of the spectrum
Parameters:
<PeakExcursion> Range: 0 to 100
Increment: 1
*RST: 5
Default unit: dB
Usage: Asynchronous command
Manual operation: See "Peak excursion" on page 192