Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 473

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
473User Manual 1802.3369.02 ─ 02
Usage: Asynchronous command
Manual operation: See "Interpolation" on page 109
ACQuire:POINts[:VALue] <RecordLength>
Sets the record length in manual record length mode. In automatic record length mode,
it sets the maximum value of the record length for automatic calculation.
The record length is the number of waveform samples that are stored in one waveform
record after processing, including interpolation.
Parameters:
<RecordLength> Range: 1000 to Device specific
Increment: 2
*RST: 1000
Default unit: pts
Usage: Asynchronous command
Manual operation: See "Record length, Record length limit" on page 108
ACQuire:POINts:ARATe?
Returns the sample rate of the ADC, before waveform processing. The result is the
interleaved sample rate or the non-interleaved one, depending on the channel usage.
Return values:
<ADCSampleRate> Range: 5 Gsample/s (interleaved), 2,5 Gsample/s (non-
interleaved)
Default unit: Hz
Usage: Query only
Asynchronous command
ACQuire:POINts:MAXimum <RecLengthLimit>
Sets the maximum record length, if ACQuire:POINts:MODE is set to AUTO.
Parameters:
<RecLengthLimit> Range: 1000 to 800E+6
Increment: 2
*RST: 10E+6
Default unit: pts
Usage: Asynchronous command
ACQuire:POINts:MODE <RecLengthMode>
Selects the mode of the waveform record length adjustment.
Parameters:
<RecLengthMode> AUTO | MANual
Acquisition and setup