Remote control commands
R&S
®
MXO 5 Series
716User Manual 1802.3369.02 ─ 02
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Enable> ON | OFF
*RST: ON
Usage: Asynchronous command
SBUS<sb>:CAN:FILTer:BIT <Frame>,<Field>,<Bit>
SBUS<sb>:CAN:FILTer:FRAMe<fr>:FLD<fl>:BIT <BitState>
Sets the operator for the data pattern in the selected field of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<BitState> ONE | ZERO
*RST: ZERO
Usage: Asynchronous command
Manual operation: See "Edit" on page 348
SBUS<sb>:CAN:FILTer:DMAX <Frame>,<Field>,<Data>
SBUS<sb>:CAN:FILTer:FRAMe<fr>:FLD<fl>:DMAX <Data_Max>
Sets the end value of a data pattern range if the operator is set to INRange or
OORANGe.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Data_Max>
Usage: Asynchronous command
Manual operation: See "Edit" on page 348
SBUS<sb>:CAN:FILTer:DMIN <Frame>,<Field>,<Data>
SBUS<sb>:CAN:FILTer:FRAMe<fr>:FLD<fl>:DMIN <Data_Min>
Specifies the data pattern, or sets the start value of a data pattern range.
Suffix:
<sb>
.
1...4, index of the serial bus