Applications
R&S
®
MXO 5 Series
237User Manual 1802.3369.02 ─ 02
Remote command:
FRANalysis:REPeat on page 608
11.1.3.2 Amplitude profile
Access: [Apps] > "General" > "FRA" > "Setup" > "Amplitude profile".
Opens a dialog to set the amplitude profile. You can then define different amplitudes for
different frequency. The amplitude profile is useful when testing sensitive circuits,
where the amplitude gets too high. In this case distortion can occur.
Step start freq, Amplitude
Set the frequency and amplitude values for the selected point.
Remote command:
FRANalysis:AMPLitude:PROFile:POINt<m>:AMPLitude on page 610
FRANalysis:AMPLitude:PROFile:POINt<m>:FREQuency on page 610
FRANalysis:AMPLitude:PROFile:COUNt on page 609
Voltage change
Selects if the voltage change is done as a single step or as a ramp.
Remote command:
FRANalysis:AMPLitude:PROFile:MODE on page 610
Add
Adds a new point to the amplitude profile.
Remote command:
FRANalysis:AMPLitude:PROFile:APOint on page 609