Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 320

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Protocol analysis
R&S
®
MXO 5 Series
320User Manual 1802.3369.02 ─ 02
"From"
Specifies the value or sets the start value of a range.
"To"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger:I2C:ADDRess on page 671
TRIGger:I2C:ADDTo on page 671
Data
Specifies the trigger conditions for the data bit pattern.
"Condition"
Sets the comparison condition to a specific value or a range.
"Value"
Specifies the value or sets the start value of a range.
Enter the bytes in msb first bit order. The maximum pattern length is
64 bit. Waveform data is compared with the pattern byte-by-byte.
Remote command:
TRIGger:I2C:DCONdition on page 672
TRIGger:I2C:DMIN on page 672
Position
Sets the number of data bytes to be skipped after the address.
Remote command:
TRIGger:I2C:DPOSition on page 672
13.3.5 I²C decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. Open the "Serial Bus" dialog for the respective bus.
2. In the "Setup" tab, enable "State".
3. In the "Display" tab, enable "Show decode table".
For a description of the display settings, see also Chapter 13.1.3, "Display",
on page 288.
The instrument captures and decodes the signal according to the standard definition
and the configuration settings.
The color-coding of the various protocol sections and errors simplifies the interpretation
of the visual display. The decode information condenses or expands, depending on the
horizontal scale. Various data formats are available to show the result values.
I²C (option R&S
MXO5-K510)
Loading ...
Loading ...