Remote control commands
R&S
®
MXO 5 Series
708User Manual 1802.3369.02 ─ 02
Manual operation: See "Data pattern" on page 355
TRIGger:CAN:DMIN <DataPattern>
Sets a data pattern, or sets the start value of a data pattern range.
Parameters:
<DataPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Data pattern" on page 355
TRIGger:CAN:FDATa:DPOSition
<DataPosition>
Defines the number of the first data byte at which the data pattern may start.
Parameters:
<DataPosition> Range: 1 to 2048
Increment: 1
*RST: 1
Usage: Asynchronous command
TRIGger:CAN:DLC <DLC>
Sets the data length code, the number of data bytes to be found. For complete defini-
tion, set also the operator with TRIGger:CAN:DLCCondition.
Parameters:
<DLC> Range: CAN: 1 to 8, CAN FD: 1 to 15 (64 bytes)
Increment: 1
*RST: 0
Usage: Asynchronous command
Manual operation: See "DLC" on page 354
TRIGger:CAN:DLCCondition <DLCOperator>
Operator to set the data length code.
The number of data bytes to be found is set with TRIGger:CAN:DLC.
Parameters:
<DLCOperator> EQUal | GETHan
For little endian transfer direction, EQUal must be set.
*RST: GETHan
Usage: Asynchronous command
Manual operation: See "DLC" on page 354