Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 478

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
478User Manual 1802.3369.02 ─ 02
USER
Probe is not detected and not known to the instrument. Set unit
and attenuation manually.
ZC10 | ZC20 | ZC30 | ZS10L | ZC03
Type of the probe
ZD01A100 | ZD01A1000
High voltage differential probes, attenuation ratio according to
the setting on the probe.
A100 = 100:1
A1000 = 1000:1
ZC02100 | ZC021000
Current probes 100 A or 1000 A according to the setting on the
probe.
*RST:
NONE
Usage: Asynchronous command
Manual operation: See "Predefined probe, name and type of the probe"
on page 120
PROBe<ch>:SETup:ATTenuation:UNIT <Unit>
Returns the unit of the connected probe if the probe is detected or predefined. For
unknown probes, you can select the required unit.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Unit> V | A | W
Voltage probe (V), current probe (A), power probe (W)
*RST: V
Usage: Asynchronous command
Manual operation: See "Vertical unit" on page 121
PROBe<ch>:SETup:BANDwidth?
Returns the bandwidth of the connected probe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<Bandwidth> Range: 10000 to 20000000000
Increment: 10
*RST: 1500000000
Default unit: Hz
Acquisition and setup
Loading ...
Loading ...