Remote control commands
R&S
®
MXO 5 Series
517User Manual 1802.3369.02 ─ 02
Usage: Asynchronous command
Manual operation: See "Trigger out" on page 157
TRIGger:ACTions:OUT:SOURce <SignalSource>
Defines when the trigger out signal is initiated: at the trigger point, when waiting for the
trigger, or when the post-trigger time is finished.
Parameters:
<SignalSource> TRIG | TRIGGER | POST | POSTTRIGGER | WAIT |
WAITTRIGGER
TRIG = TRIGGER, POST = POSTTRIGGER, WAIT = WAIT-
TRIGGER
*RST: TRIG
Usage: Asynchronous command
Manual operation: See "Signal source" on page 158
TRIGger:ACTions:OUT:DELay <Delay>
Defines the delay of the first pulse edge to the trigger point. The minimum delay is
600 ns.
Parameters:
<Delay> Range: 8E-07 to 1
Increment: 4E-09
*RST: 8E-07
Default unit: s
Usage: Asynchronous command
Manual operation: See "Delay" on page 158
TRIGger:ACTions:OUT:PLENgth <PulseLength>
Sets the length of the trigger out pulse.
Parameters:
<PulseLength> Range: 1.6E-08 to 0.05
Increment: 1.6E-08
*RST: 9.6E-08
Default unit: s
Usage: Asynchronous command
Manual operation: See "Pulse length" on page 158
TRIGger:ACTions:OUT:POLarity <Polarity>
Sets the polarity of the trigger out pulse, which is the direction of the first pulse edge.
Trigger