Remote control commands
R&S
®
MXO 5 Series
753User Manual 1802.3369.02 ─ 02
<n> 0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Skew> Range: -2E-07 to 2E-07
Increment: 2E-10
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "D0-D15" on page 384
PBUS<pb>:BIT<n>[:STATe] <Assigned>
Enables the selected logic group. The corresponding signal icon appears on the signal
bar.
If another active bus already uses the selected digital channel, the instrument disables
the other bus to avoid conflicts.
Suffix:
<pb>
.
1...4, index of the logic group
<n> 0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Assigned> ON | OFF
*RST: OFF
Usage: Asynchronous command
PBUS<pb>:CLEar
Removes all assigned digital channels from the bus.
Suffix:
<pb>
.
1...4, index of the logic group
Usage: Setting only
Asynchronous command
PBUS<pb>:CLOCk <ClockSource>
Selects the digital channel used as clock.
Suffix:
<pb>
.
1 | 2
Selects the logic. The clocked bus is available on Logic 1 and 2.
Mixed signal option (option R&S MXO5-B1)