Remote control commands
R&S
®
MXO 5 Series
506User Manual 1802.3369.02 ─ 02
17.9.8 Interval trigger
TRIGger:EVENt<ev>:INTerval:DELTa...............................................................................506
TRIGger:EVENt<ev>:INTerval:RANGe............................................................................. 506
TRIGger:EVENt<ev>:INTerval:SLOPe..............................................................................507
TRIGger:EVENt<ev>:INTerval:WIDTh.............................................................................. 507
TRIGger:EVENt<ev>:INTerval:DELTa <WidthDelta>
Sets a range around the interval width value specified with TRIGger:EVENt<ev>:
INTerval:WIDTh.
Suffix:
<ev>
.
1, can be omitted
Parameters:
<WidthDelta> Range: 0 to 10
Increment: 1E-07
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "±Delta" on page 150
TRIGger:EVENt<ev>:INTerval:RANGe <RangeMode>
Defines the range of an interval in relation to the interval width specified using
TRIGger:EVENt<ev>:INTerval:WIDTh and TRIGger:EVENt<ev>:INTerval:
DELTa.
Suffix:
<ev>
.
1, can be omitted
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers on pulses inside a given range. The range is defined by
the interval width ±delta.
OUTSide
Triggers on pulses outside a given range. The range is defined
by the interval width ±delta.
SHORter
Triggers on pulses shorter than the given interval width.
LONGer
Triggers on pulses longer than the given interval width.
*RST:
OUTSide
Usage: Asynchronous command
Manual operation: See "Range" on page 149