Remote control commands
R&S
®
MXO 5 Series
467User Manual 1802.3369.02 ─ 02
DCLimit
Connection with 1 MΩ termination, passes both DC and AC
components of the signal.
AC
Connection with 1 MΩ termination through DC capacitor,
removes DC and very low-frequency components. The wave-
form is centered on zero volts.
Usage: Asynchronous command
Manual operation: See "Coupling" on page 116
CHANnel<ch>:INVert
<InvertChannel>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the volt-
age values of all signal components against the ground level.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<InvertChannel> ON | OFF
ON: inverted waveform
OFF: normal waveform
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Invert channel" on page 116
CHANnel<ch>:SKEW:TIME <Offset>
Sets a skew value to compensate for the delay of the measurement setup or from the
circuit specifics that the instrument cannot compensate automatically. It affects only the
selected input channel.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Offset> Range: -1E-07 to 1E-07
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Skew" on page 117