Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 229

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Spectrum analysis
R&S
®
MXO 5 Series
229User Manual 1802.3369.02 ─ 02
"Peak on waveform" > "Off" "Peak on waveform" > "On"
Remote command:
CALCulate:SPECtrum<sp>:PLISt:SPOW on page 599
Threshold
Sets an absolute threshold as an additional condition for the peak search. Only peaks
that exceed the threshold are detected.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Remote command:
CALCulate:SPECtrum<sp>:THReshold on page 594
CURSor<cu>:THReshold on page 587
Peak excursion
Defines a relative threshold, the minimum level value by which the waveform must rise
by to be considered as a peak. For the peak to be considered, the excursion threshold
must be met both on the left and the right sides of the peak.
To avoid identifying noise peaks, enter a peak excursion value that is higher than the
noise levels.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Example:
In the figure below, 2 peaks are shown, that rise above the defined "Threshold". Peak 1
fulfills the defined "Peak excursion" value and is counted as a peak. Peak 2 does not
fulfill the defined "Peak excursion" value on the right and is not counted as a peak.
Spectrum peak list
Loading ...
Loading ...
Loading ...