Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 593

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
593User Manual 1802.3369.02 ─ 02
Suffix:
<sp>
.
1...4, index of the spectrum
Parameters:
<VerticalRange> Range: 0 to 1000000000000000
Increment: 0.01
*RST: 100
Usage: Asynchronous command
Manual operation: See "Vertical range" on page 226
CALCulate:SPECtrum<sp>:MAGNitude:SCALe
<Unit>
Sets the unit for the y-axis.
Suffix:
<sp>
.
1...4, index of the spectrum
Parameters:
<Unit> LINear | DB | DBM | DBV | DBUV | DBS | DBHZ
*RST: DBM
Usage: Asynchronous command
Manual operation: See "Magnitude unit" on page 225
CALCulate:SPECtrum<sp>:PRESet
Presets the spectrum measurement.
Suffix:
<sp>
.
1...4, index of the spectrum
Usage: Setting only
Asynchronous command
Manual operation: See "Spectrum preset" on page 225
CALCulate:SPECtrum<sp>:SOURce <Source>
Selects the source for the spectrum.
Suffix:
<sp>
.
1...4, index of the spectrum
Parameters:
<Source> C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 |
M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8
Usage: Asynchronous command
Manual operation: See "Source" on page 222
Spectrum analysis
Loading ...
Loading ...
Loading ...