Remote control commands
R&S
®
MXO 5 Series
504User Manual 1802.3369.02 ─ 02
Manual operation: See "Vertical condition" on page 146
TRIGger:EVENt<ev>:WINDow:TIME <TimeRangeMode>
Available for TRIGger:EVENt<ev>:WIDTh:RANGe = WITHin and OUTSide.
Selects how the time limit of the window is defined.
You can specify the width with TRIGger:EVENt<ev>:WIDTh:WIDTh and the delta
with TRIGger:EVENt<ev>:WINDow:DELTa.
Suffix:
<ev>
.
1, can be omitted
Parameters:
<TimeRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time Width - Delta and for Width + Delta at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than Width - Delta or longer than Width +
Delta.
SHORter
Triggers if the signal crosses vertical limits before the specified
width time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
width time is reached.
*RST:
WITHin
Usage: Asynchronous command
Manual operation: See "Time condition" on page 147
TRIGger:EVENt<ev>:WINDow:WIDTh <Width>
Depending on the values of TRIGger:EVENt<ev>:WINDow:RANGe the width sets:
For the ranges Within and Outside, the width defines the center of a time range.
The range is defined by the limits "±Delta", see TRIGger:EVENt<ev>:WINDow:
DELTa.
For the ranges Shorter and Longer, it defines the maximum and minimum time
lapse, respectively.
Suffix:
<ev>
.
1, can be omitted