Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 651

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
651User Manual 1802.3369.02 ─ 02
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Threshold>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 296
SBUS<sb>:SPI:MISO:HYSTeresis <Hysteresis>
Sets a value for the hysteresis for the MISO channel.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Hysteresis>
Usage: Asynchronous command
Manual operation: See "Threshold" on page 296
SBUS<sb>:SPI:MISO:POLarity <MISOPolarity>
Selects if the transmitted signal for the respective line is active high (high = 1) or active
low (low = 1).
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<MISOPolarity> ACTLow | ACTHigh
*RST: ACTHigh
Usage: Asynchronous command
Manual operation: See "Polarity: MOSI, MISO, CS" on page 296
SBUS<sb>:SPI:MISO:SOURce MISOSource
Sets the input channel of the MISO line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
MISOSource C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5
| D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2
| M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 |
R8
Usage: Asynchronous command
Manual operation: See "SCLK,MOSI,MISO,CS" on page 295
Protocols
Loading ...
Loading ...
Loading ...