Remote control commands
R&S
®
MXO 5 Series
776User Manual 1802.3369.02 ─ 02
Suffix:
<wg>
.
1..2
Usage: Setting only
Asynchronous command
Manual operation: See "Open" on page 401
WGENerator<wg>:ARBGen:RUNMode <RunMode>
Sets the duration for which the signal of the arbitrary generator will be output after the
trigger event.
Suffix:
<wg>
.
1..2
Parameters:
<RunMode> SINGle | REPetitive
*RST: REPetitive
Usage: Asynchronous command
Manual operation: See "Run mode" on page 400
WGENerator<wg>:ARBGen:RUNSingle
Executes a single period of the arbitrary signal generator, if WGENerator<wg>:
ARBGen:RUNMode is set to SINGle.
Suffix:
<wg>
.
1..2
Usage: Event
Asynchronous command
Manual operation: See "Run single" on page 400
WGENerator<wg>:ARBGen:SAMPles?
Sets the sample rate for the arbitrary waveform.
Suffix:
<wg>
.
1..2
Return values:
<NumSamples> Range: 0 to 128000000
Increment: 10
*RST: 0
Default unit: pts
Usage: Query only
Asynchronous command
Manual operation: See "Number of samples" on page 401