Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 567

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
567User Manual 1802.3369.02 ─ 02
MEASurement<mg>:STATistics[:ENABle] <GlobalEnable>
Enables statistics calculation for all measurements.
Suffix:
<mg>
.
Irrelevant, omit the suffix.
Parameters:
<GlobalEnable> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Statistics" on page 212
MEASurement<mg>:MULTiple <GlobalMeassAll>
If ON, the measurement is performed repeatedly if the measured parameter occurs
several times inside the acquisition or defined gate. All results are included in evalua-
tion, e.g. in statistics. To set the number of results to be considered, use
MEASurement<mg>:MNOMeas.
Suffix:
<mg>
.
Irrelevant, omit the suffix.
Parameters:
<GlobalMeassAll> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Measure all events" on page 212
MEASurement<mg>:MNOMeas <MaxMeasPerAcq>
Sets the maximum number of measurements per acquisition if MEASurement<mg>:
MULTiple is on. The setting affects all measurements.
Suffix:
<mg>
.
Irrelevant, omit the suffix.
Parameters:
<MaxMeasPerAcq> Range: 2 to 1E6
Increment: 1
*RST: 1E6
Usage: Asynchronous command
Manual operation: See "Max. No. of events" on page 213
Automatic measurements
Loading ...
Loading ...