Network operation and remote control
R&S
®
MXO 5 Series
424User Manual 1802.3369.02 ─ 02
Bit No. Meaning
5
WTRIgger
Available only for single acquisitions or the first acquisition of continuous acquisitions.
The wait for trigger status bit indicates that the instrument is ready to trigger, and the pre-trigger
time is expired.
The bit is set for as long as a measurement is running.
6
TRIggered
Available only for single acquisitions or the first acquisition of continuous acquisitions.
The triggered status bit indicates that the instrument has been triggered.
7 - 15 Not used
16.5.3.4 STATus:QUEStionable register
This register contains specific information on instrument operation and signal reliability.
It can be read using the commands STATus:QUEStionable:CONDition? and
STATus:QUEStionable[:EVENt]?.
The remote commands for the STATus:QUEStionable register are described in Chap-
ter 17.19, "Status reporting", on page 780.
Table 16-5: Bits in the STATus:QUEStionable register
Bit
No.
Meaning
0 - For future use.
1
PPSupply
This bit is set if probe power supply overload occurs.
2
POVerload
This bit is set if a probe overload occurs.
3
COVerload
This bit is set if a questionable channel overload occurs.
4
TEMPerature
This bit is set if a questionable temperature occurs.
5
ADCState
The bit is set if the signal is clipped on the upper or lower edge of the screen over-
flow of the ADC occurs.
6
GOVerload
This bit is set if a generator overload occurs.
7 - For future use.
8
NOALigndata
This bit is set if no alignment data is available - the instrument is uncalibrated.
9
LIMit
This bit is set if a limit value is violated.
10
MARGin
This bit is set if a margin value is violated, and the limit value is not violated.
11
IMPRecise
This bit is set if the magnitude of the signal is too low to get reliable measurement
results.
12 - For future use.