Remote control commands
R&S
®
MXO 5 Series
607User Manual 1802.3369.02 ─ 02
FRANalysis:MEASurement:RBW <RBW>
Sets the resolution bandwidth, which determines the number of measurements that are
used for creating the plot.
Parameters:
<RBW> HIGH | MID | LOW
*RST: MID
Usage: Asynchronous command
Manual operation: See "RBW" on page 239
FRANalysis:OUTPut[:SOURce] <OutputChannel>
Sets the channel for the output signal of the DUT.
Parameters:
<OutputChannel> C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8
*RST: C2
Usage: Asynchronous command
Manual operation: See "Output" on page 235
FRANalysis:POINts:LOGarithmic <PtsPerDecade>
Selects the number of points that are measured per decade, if FRANalysis:POINts:
MODE is set to DECade.
Parameters:
<PtsPerDecade> Range: 10 to 500
Increment: 1
*RST: 10
Usage: Asynchronous command
Manual operation: See "Points" on page 236
FRANalysis:POINts:MODE <PointsMode>
Selects, if the number of points for the FRA are measured as total or per decade.
You can set the number of points with FRANalysis:POINts:TOTal/ FRANalysis:
POINts:LOGarithmic.
Parameters:
<PointsMode> TOTal | DECade
*RST: DECade
Usage: Asynchronous command
Manual operation: See "Points" on page 236