Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 147

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Trigger
R&S
®
MXO 5 Series
147User Manual 1802.3369.02 ─ 02
Lower level
Sets the lower voltage limit.
Remote command:
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer on page 502
Time condition
Available for "Vertical condition" = "Stay within"/"Stay outside".
Selects how the time limit of the window is defined.
"Within"
Triggers if the signal stays inside or outside the vertical window limits
at least for the time Width - Delta and for Width + Delta at the most.
"Outside"
"Outside" is the opposite definition of "Within". The instrument triggers
if the signal stays inside or outside the vertical window limits for a
time shorter than Width - Delta or longer than Width + Delta.
"Shorter"
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
"Longer"
Triggers if the signal crosses vertical limits after the specified "Width"
time is reached.
Remote command:
TRIGger:EVENt<ev>:WINDow:TIME on page 504
Width
For the ranges "Within" and "Outside", the width defines the center of a time range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
Remote command:
TRIGger:EVENt<ev>:WINDow:WIDTh on page 504
±Delta
Defines a range around the width value.
Remote command:
TRIGger:EVENt<ev>:WINDow:DELTa on page 503
7.3.7 Timeout trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Timeout"
The timeout trigger checks if the signal stays above or below the threshold voltage for
a specified time lapse. In other words, the trigger occurs if the trigger source does not
have the expected transition within the specified time.
Trigger types
Loading ...
Loading ...
Loading ...