Protocol analysis
R&S
®
MXO 5 Series
302User Manual 1802.3369.02 ─ 02
Type
Selects the trigger type for SPI analysis.
Some trigger types have additional settings that can be defined. In this case, the "Set
details" button appears next to "Type" the function. Open it for a detailed definition of
the trigger conditions.
"Start of frame"
Sets the trigger to the beginning of the frame.
"End of frame"
Sets the trigger to the end of the frame.
"MOSI"
Sets the trigger to a specified data pattern expected on the MOSI
line.
Available, if there is a channel assigned to the "MOSI" line in the
"Setup" tab.
"MISO"
Sets the trigger to a specified data pattern expected on the MISO
line.
Available, if there is a channel assigned to the "MISO" line in the
"Setup" tab.
Remote command:
TRIGger:SPI:TYPE on page 660
Data conditions
The trigger on MOSI and MISO patterns is defined in the same way:
SPI bus (option R&S MXO5-K510)