Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 261

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Applications
R&S
®
MXO 5 Series
261User Manual 1802.3369.02 ─ 02
10/50/90
20/50/80
User defined: Enter "Upper level", "Middle level", and "Lower level".
For example, for "5/50/95" the levels are set to the following values:
Lower level = 5% of the signal amplitude
Middle level = 50% of the signal amplitude
Upper level = 95% of the signal amplitude
Remote command:
POWer<m>:HARMonics:REFLevel<rl>:RELative:MODE on page 638
POWer<m>:QUALity:REFLevel<rl>:RELative:MODE on page 638
Upper level, Middle level,Lower level ← Relative Level mode
Define the reference levels in percent, if "Level mode" is set to "User-defined".
Remote command:
POWer<m>:HARMonics:REFLevel<rl>:RELative:LOWer on page 638
POWer<m>:HARMonics:REFLevel<rl>:RELative:MIDDle on page 638
POWer<m>:HARMonics:REFLevel<rl>:RELative:UPPer on page 639
POWer<m>:QUALity:REFLevel<rl>:RELative:LOWer on page 638
POWer<m>:QUALity:REFLevel<rl>:RELative:MIDDle on page 638
POWer<m>:QUALity:REFLevel<rl>:RELative:UPPer on page 639
Hysteresis
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Remote command:
POWer<m>:HARMonics:REFLevel<rl>:RELative:HYSTeresis on page 637
POWer<m>:HARMonics:REFLevel<rl>:ABSolute:HYSTeresis on page 635
POWer<m>:QUALity:REFLevel<rl>:RELative:HYSTeresis on page 637
POWer<m>:QUALity:REFLevel<rl>:ABSolute:HYSTeresis on page 635
11.2.4 Deskew
In the dialog box you can perform a deskew of your current and voltage probes.
Required equipment:
R&S RT-ZF20 power deskew fixture
Rohde & Schwarz voltage probe
Rohde & Schwarz current probe
Power analysis (option R&S
MXO5-K31)