Remote control commands
R&S
®
MXO 5 Series
480User Manual 1802.3369.02 ─ 02
PROBe<ch>:SETup:NAME?
Queries the name of the probe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<Name> Name string
Usage: Query only
Asynchronous command
Manual operation: See "Predefined probe, name and type of the probe"
on page 120
PROBe<ch>:SETup:STATe?
Queries if the probe at the specified input channel is active (detected) or not active (not
detected). To switch the probe on, use CHANnel<ch>:STATe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<State> DETected | NDETected
*RST: NDETected
Usage: Query only
Asynchronous command
Manual operation: See "Predefined probe, name and type of the probe"
on page 120
PROBe<ch>:SETup:TYPE?
Queries the class of the probe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Return values:
<Type> String containing the probe class, e.g. None (no probe detec-
ted), passive, current, active diff., active single-ended.
Usage: Query only
Asynchronous command
Manual operation: See "Predefined probe, name and type of the probe"
on page 120