Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 289

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Protocol analysis
R&S
®
MXO 5 Series
289User Manual 1802.3369.02 ─ 02
Remote command:
SBUS<sb>:RESult on page 645
Decode layer
Selects the decode layer.
Decoding is performed in several steps, and the results are presented in the decode
table. The decode layer selects an interim step for which the decoding result is shown
in the honeycomb display.
The available values are protocol-specific.
"Off"
No decode layer is displayed.
"Edges"
All edges.
"Bits"
All bits.
"Words"
All words.
Data format
Sets the data format for the values displayed in the decode table and in the combs of
the decoded signal.
Available formats are: hex, octal, binary, ASCii, signed, unsigned.
For more information on the available formats, see Chapter 13.1.8, "Bit pattern editor",
on page 292.
Remote command:
SBUS<sb>:FORMat on page 645
Zoom coupling
If enabled, the decode zoom and result table are synchronized. If you select a row in
the result table, this result is shown in the decode zoom.
Remote command:
SBUS<m>:ZCOupling on page 646
Symbols
Symbol lists are protocol-specific. They are described in the respective configuration
chapter of the protocol.
13.1.4 Filter
In the "Filter" tab, you can define the settings to display only the frames that match the
selected filter conditions.
The settings are protocol specific and are described in the corresponding protocol
chapter.
13.1.5 Trigger
The "Trigger" tab gives quick access to the "Trigger" dialog.