Remote control commands
R&S
®
MXO 5 Series
488User Manual 1802.3369.02 ─ 02
Parameters:
<Gain> Range: 0.0001 to 10000
Increment: 0.0001
*RST: 1
Default unit: V/V
Usage: Asynchronous command
Manual operation: See "Gain, Manual gain" on page 130
PROBe<ch>:SETup:DEGauss
Demagnetizes the core if it has been magnetized by switching the power on and off, or
by an excessive input. Always carry out demagnetizing before measurement.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Usage: Event
Asynchronous command
Manual operation: See "DeGauss" on page 131
PROBe<ch>:SETup:OFFSet:ZADJust <ZeroAdjustValue>
Set the waveform to zero position. It corrects the effect of a voltage offset or tempera-
ture drift. To set the value by the instrument, use PROBe<ch>:SETup:OFFSet:AZERo
.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<ZroAdjVal> Range: -100 to 100
Increment: 0.1
*RST: 0
Default unit: %
Usage: Asynchronous command
Manual operation: See "Zero adjust" on page 131
PROBe<ch>:SETup:OFFSet:STPRobe
Saves the zero adjust value in the probe box. If you connect the probe to another
channel or to another Rohde & Schwarz oscilloscope, the value is read out again, and
you can use the probe without further adjustment.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Acquisition and setup