Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 512

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
512User Manual 1802.3369.02 ─ 02
Return values:
<AutoTime> Range: 1E-07 to 10
Increment: 0.0002
*RST: 0.001
Default unit: s
Usage: Query only
Asynchronous command
Manual operation: See "Holdoff mode" on page 154
TRIGger:HOLDoff:SCALing <AutoTimeScl>
Sets the auto time scaling factor that the horizontal scale is multipied with, if TRIGger:
HOLDoff:MODETRIGger:HOLDoff:MODETRIGger:HOLDoff:MODE is set to AUTO.
Auto time = Auto time scaling * Horizontal scale
The next trigger occurs only after this time has passed.
Parameters:
<AutoTimeScl> Range: 0.001 to 1000
Increment: 1
*RST: 0.5
Usage: Asynchronous command
Manual operation: See "Holdoff mode" on page 154
TRIGger:HOLDoff:EVENts <Events>
Defines the number of triggers to be skipped, if TRIGger:HOLDoff:MODETRIGger:
HOLDoff:MODETRIGger:HOLDoff:MODE is set to EVENts. The next trigger only
occurs when this number of events is reached.
Parameters:
<Events> Range: 1 to 2147483647
Increment: 10
*RST: 1
Usage: Asynchronous command
Manual operation: See "Holdoff mode" on page 154
TRIGger:HOLDoff:MAX <RandomMaxTime>
Defines the upper limit for the random time holdoff, if TRIGger:HOLDoff:MODE
TRIGger:HOLDoff:MODETRIGger:HOLDoff:MODE is set to RANDom.
Parameters:
<RandomMaxTime> Range: 1E-07 to 10
Increment: 0.0002
*RST: 0.002
Default unit: s
Trigger
Loading ...
Loading ...
Loading ...