Trigger
R&S
®
MXO 5 Series
145User Manual 1802.3369.02 ─ 02
Remote command:
TRIGger:EVENt<ev>:LEVel<n>:RUNT:LOWer on page 499
Range
Selects how the time limit of the runt pulse is defined.
"Any runt"
Triggers on all runts fulfilling the level condition, without time limita-
tion.
"Longer"
Triggers on runts longer than the given "Runt width".
"Shorter"
Triggers on runts shorter than the given "Runt width".
"Within"
Triggers if the runt length is inside a given time range. The range is
defined by "Runt width" and "±Delta".
"Outside"
Triggers if the runt length is outside a given time range. The range
definition is the same as for "Within" range.
Remote command:
TRIGger:EVENt<ev>:RUNT:RANGe on page 501
Runt width
For the ranges "Shorter" and "Longer", the runt width defines the maximum and mini-
mum pulse width, respectively.
For the ranges "Within" and "Outside", the runt width defines the center of a range
which is defined by "±Delta".
Remote command:
TRIGger:EVENt<ev>:RUNT:WIDTh on page 501
±Delta
Defines a range around the runt width value.
Remote command:
TRIGger:EVENt<ev>:RUNT:DELTa on page 500
7.3.6 Window trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Window"
The window trigger checks the signal run in relation to a "window". The window is
formed by the upper and lower voltage levels. The trigger condition is fulfilled, if the
waveform enters or leaves the window, or if the waveform stays inside or outside for a
time longer or shorter than specified.
With the window trigger, you can display longer transient effects.