Remote control commands
R&S
®
MXO 5 Series
528User Manual 1802.3369.02 ─ 02
Operation <Expression> Comment
Rescale "Rescale(C1,a,b)" a = scale, default = 1
b = offset, default = 0
FIR "FIR(Type,C1,Cut-Off,Characteristics)"
Examples:
"FIR(highpass,C1,10000000,Gaussian)"
"FIR(lowpass,C1,10000000,rectangle)"
Type = lowpass, highpass
Cut-Off = limit frequency
Characteristics = Gaussian, rectangle
Cut-Off can get any value between
4 GHz and 1 kHz
Suffix:
<m>
.
1...8, index of the math waveform
Parameters:
<Expression> String with regular expression for calculation
Example:
CALC:MATH 'C1*C2'
Defines the multiplication of waveforms channel 1 and channel
2.
Usage: Asynchronous command
Manual operation: See "Operator" on page 167
CALCulate:MATH<m>:VERTical:SCALe[:VALue] <VerticalScale>
Sets the scale of the y-axis in the math function diagram. The value is defined as
"<unit> per division", e.g. 50
mV/div. In this case, the horizontal grid lines are displayed
in intervals of 50 mV.
Suffix:
<m>
.
1...8, index of the math waveform
Parameters:
<VerticalScale> Range: 1E-12 to 100E+12
Increment: 1E-05
*RST: 0.5
Default unit: V
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 39
CALCulate:MATH<m>:VERTical:OFFSet <VerticalOffset>
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Negative values move the waveform up, positive values move it down.
Suffix:
<m>
.
1...8, index of the math waveform
Waveform analysis