Waveform analysis
R&S
®
MXO 5 Series
169User Manual 1802.3369.02 ─ 02
Where f_HP is the requested highpass limit frequency and f_LP the equivalent lowpass
frequency that has to comply with the limits given above.
Characteristics
Selects if a Gaussian or a rectangular shape is used for the lowpass filter. The high-
pass filter is always Gaussian.
8.2.4 Math scale settings
Access:[Math] > "Scale" tab
Scale mode
By default, the vertical scale is adapted to the current measurement results automati-
cally to provide an optimal display. However, if necessary, you can define scaling val-
ues manually to suit your requirements.
"Manual"
Enter the required values for "Vertical scale" and "Vertical offset".
"Auto"
"Vertical scale" and "Vertical offset" are read-only.
Remote command:
CALCulate:MATH<m>:VERTical:SCALe:MODE on page 529
Vertical scale
Sets the scale of the y-axis in the math function diagram. The value is defined as
"<unit> per division", e.g. 50
mV/div. In this case, the horizontal grid lines are displayed
in intervals of 50 mV.
If Scale mode is set to "Auto", this setting is read-only.
Remote command:
CALCulate:MATH<m>:VERTical:SCALe[:VALue] on page 528
Vertical offset
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Negative values move the waveform up, positive values move it down.
If Scale mode is set to "Auto", this setting is read-only.
Mathematics