Rohde & Schwarz MXO58-200 Rohde & Schwartz oscilloscope

Rohde & Schwartz MXO5 user manual - Page 469

For MXO58-200.

PDF File Manual, 824 pages, Read Online | Download pdf file

MXO58-200 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 5 Series
469User Manual 1802.3369.02 ─ 02
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<ExtAtt> Values depend on the selected scale (CHANnel<ch>:
EATScale on page 468) and the unit of the waveform. See
"External Attenuation: Scale, Attenuation" on page 119. Limits
below are for linear scale.
Range: 0.005 to 1000000
Increment: 0.01
*RST: 1
Usage: Asynchronous command
Manual operation: See "External Attenuation: Scale, Attenuation" on page 119
CHANnel<ch>:IMPedance
<Impedance>
Sets the impedance of the connected probe for power calculations and measurements.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Impedance> Range: 0.1 to 100000
Increment: 1
*RST: 50
Default unit:
Usage: Asynchronous command
Manual operation: See "Impedance" on page 119
17.8.5 Waveform data export
To set the export data format, see FORMat[:DATA].
CHANnel<ch>[:WAVeform<wf>]:DATA:HEADer?............................................................... 469
CHANnel<ch>[:WAVeform<wf>]:DATA[:VALues]?.............................................................. 470
CHANnel<ch>[:WAVeform<wf>]:DATA:HEADer?
Returns the header of channel waveform data. The header contains attributes of the
waveform.
Table 17-2: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
Acquisition and setup