Measurements
R&S
®
MXO 5 Series
208User Manual 1802.3369.02 ─ 02
Remote command:
MEASurement<mg>:GATE on page 563
Label
Adds a user-defined label to the measurement.
Envelope
This setting is only available for measurements on envelope waveforms, see Acquisi-
tion mode.
"Both"
The upper and the lower envelope are used in measurements. For
time measurements, the averages of min and max values are used,
that is, the measurement is performed on the average waveform built
from the upper and lower envelope.
"Maximum"
Measurements are performed on the upper envelope.
"Minimum"
Measurements are performed on the lower envelope.
Remote command:
MEASurement<mg>:ENVSelect on page 559
Pulse count
Available, if "Specific type" is set to "Pulse train".
Sets the number N of positive pulses for the "Pulse train" measurement. This measure-
ment measures the duration of N positive pulses from the rising edge of the first pulse
to the falling edge of the N-th pulse.
Remote command:
MEASurement<mg>:AMPTime:PTCount on page 561
Pulse slope
Available, if "Specific type" is set to "Pulse count".
Sets the first slope of the pulses to be counted.
"Positive"
Positive pulses are counted.
"Negative"
Negative pulses are counted.
"Either"
Both positive and negative pulses are counted.
Remote command:
MEASurement<mg>:AMPTime:PSLope on page 561
Edges slope
Available, if "Specific type" is set to "Edge count" or "Delay to trigger".
Sets the edge direction to be used for delay measurement: positive, negative, or either
edge.
Remote command:
MEASurement<mg>:AMPTime:ESLope on page 560
Automatic measurements