Remote control commands
R&S
®
MXO 5 Series
463User Manual 1802.3369.02 ─ 02
17.8.3 Roll mode
TIMebase:ROLL:ENABle................................................................................................ 463
TIMebase:ROLL:MTIMe..................................................................................................463
ACQuire:ROLLmode:OSCapture..................................................................................... 463
ACQuire:ROLLmode:POINts?......................................................................................... 464
TIMebase:ROLL:STATe?.................................................................................................464
TIMebase:ROLL:ENABle <Mode>
Selects, if the roll mode is set automatically by the instrument or if it is turned off.
Parameters:
<Mode> AUTO | OFF
*RST: AUTO
Usage: Asynchronous command
Manual operation: See "Mode" on page 103
TIMebase:ROLL:MTIMe <MinAcquTime>
Sets the minimum acquisition time for automatic start of the roll mode.
Parameters:
<MinAcquTime> Range: 0.5 to 100000
Increment: 1
*RST: 2
Default unit: s
Usage: Asynchronous command
Manual operation: See "Start roll time" on page 103
ACQuire:ROLLmode:OSCapture <OffScreenCapture>
If enabled, the analyzable waveform in roll mode is extended. Samples that arrive the
left edge of the display are kept, they are moved to the unvisible area on the left.
Parameters:
<OffScreenCapture> ON | OFF
*RST: OFF
Example: See Chapter 17.4.3, "Data transfer in roll mode", on page 435.
Usage: Asynchronous command
Manual operation: See "Off screen capture" on page 103