![](https://files.manualsfile.com/11202903-mxo54-2000/bgb1.png)
Waveform analysis
R&S
®
MXO 5 Series
177User Manual 1802.3369.02 ─ 02
5. To load the waveform from the specified file, tap "Open".
The selected waveform is loaded as the specified reference waveform.
8.4.2 Settings for reference waveforms
To compare waveforms and analyze differences between waveforms, you can use up
to eight reference waveforms R1 to R8.
Each reference waveform has its own memory on the instrument. You can also save
an unlimited number of reference waveforms and load them for further use.
The display of a reference waveform is independent from the display of the source
waveform; you can move, stretch and compress the curve vertically and horizontally.
8.4.2.1 Reference waveform setup
Access: [Ref] key > "Setup" tab
Reference waveforms