Remote control commands
R&S
®
MXO 5 Series
610User Manual 1802.3369.02 ─ 02
FRANalysis:AMPLitude:PROFile:MODE <VoltageChange>
Selects if the voltage change is done as a single step or as a ramp.
Parameters:
<VoltageChange> SINGle | RAMP
*RST: SINGle
Usage: Asynchronous command
Manual operation: See "Voltage change" on page 237
FRANalysis:AMPLitude:PROFile:POINt<m>:FREQuency
<Frequency>
Sets the start frequency for the selected point.
Suffix:
<m>
.
Index of the amplitude profile point
Parameters:
<Frequency> Range: 0.01 to 100E6
Increment: 1
*RST: 100
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Step start freq, Amplitude" on page 237
FRANalysis:AMPLitude:PROFile:POINt<m>:AMPLitude <Amplitude>
Sets the amplitude for the selected point.
Suffix:
<m>
.
Index of the amplitude profile point
Parameters:
<Amplitude> Range: 0 to 5.995
Increment: 0.1
*RST: 1
Default unit: Vpp
Usage: Asynchronous command
Manual operation: See "Step start freq, Amplitude" on page 237
17.15.1.3 Frequency response analysis calibration settings
FRANalysis:CALibration:CALibration................................................................................611
FRANalysis:CALibration:RESult?.....................................................................................611
FRANalysis:CALibration:STATe........................................................................................611
Applications