Remote control commands
R&S
®
MXO 5 Series
465User Manual 1802.3369.02 ─ 02
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Show channel" on page 115
CHANnel<ch>:SCALe <Scale>
Sets the vertical scale, which defines the displayed amplitude of the selected wave-
form.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Scale> Range: 0.001 to 1
Increment: 0.001
*RST: 0.05
Default unit: Depends on the connected probe
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 39
CHANnel<ch>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. The command is
an alternative to CHANnel<ch>:SCALe.
Suffix:
<ch>
.
1 to 8, depending on instrument model. Index of the analog
channel.
Parameters:
<Range> Range: 0.01 to 10
Increment: 0.01
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
CHANnel<ch>:OFFSet <Offset>
Sets the offset voltage, which corrects an offset-affected signal. The vertical center of
the selected channel is shifted by the offset value and the signal is repositioned within
the diagram.
Acquisition and setup
Loading ...
Loading ...