Measurements
R&S
®
MXO 5 Series
193User Manual 1802.3369.02 ─ 02
Example:
In the figure below, 2 peaks are shown, that rise above the defined "Threshold". Peak 1
fulfills the defined "Peak excursion" value and is counted as a peak. Peak 2 does not
fulfill the defined "Peak excursion" value on the right and is not counted as a peak.
Threshold
Peak excursion
1 2
Remote command:
CALCulate:SPECtrum<sp>:PEXCursion on page 594
CURSor<cu>:PEXCursion on page 587
9.2 Automatic measurements
You can perform up to 24 different measurements simultaneously.
There are two methods to start a measurement, each with slightly different effects:
Using the "Measure" icon on the toolbar:
See: "To start a measurement using the toolbar icon" on page 193.
Pressing the [Measure] key on the front panel, or using the "Measurement" menu.
See: "To configure measurements in the Measurement dialog" on page 194.
To start a measurement using the toolbar icon
1. Tap the "Measure" icon on the toolbar.
The measurement overlay menu opens.