Network operation and remote control
R&S
®
MXO 5 Series
414User Manual 1802.3369.02 ─ 02
Download files from the instrument to the operating computer. Multiple files selec-
tion is possible.
Preview file content of text files with maximum size of 500 kbyte.
After attaching a USB flash device, creating a screenshot or other actions, the file man-
ager might not display the last change:
Click the "Refresh" icon on the right of the address field to update the display.
Note that the "Eject" button for USB flash devices has no function when operating the
instrument remotely.
16.2.2.4 Mapping the instrument's file system as network drive
Using web-based distributed authoring and versioning (WebDAV), you can access and
exchange files over the internet: move, cut, copy and paste files from the instrument to
the computer and vice versa. Therefore, map the instrument's file system as network
drive in the file explorer of the operating computer:
The instrument is connected to LAN, and the connection status icon in the upper right
corner of the instrument's display is green.
1. Open the file explorer on the operating computer.
2. In the left list, right-click "Network" or the entry of the operating computer, and
select "Map network drive".
3. Enter the path to the instrument's file system:
http://<hostname>/webdavData.
Example: http://MXO58-123456/webdavData
You can find the hostname in the "Device summary", which is shown when you tap
the Rohde & Schwarz logo in the upper right corner of the instrument's screen. The
default host name is <instrument_model>-<serial_number>.
4. Activate "Connect using different credentials".
5. Click "Finish".