[Run / Stop]
Starts and stops the continuous acquisition. The [Run / Stop] key lights green when the
acquisition is running. When the acquisition is stopped, the key lights red.
[Single]
Starts a defined number of acquisitions. The [Single] key lights green when the acquisi-
tion is running. When the acquisition is stopped, the key lights red.
[Auto Norm]
Toggles the trigger mode between "Auto" and "Normal". The current setting is shown
on the trigger label.
Auto
The instrument triggers repeatedly after a time interval if the trigger
conditions are not fulfilled. If a real trigger occurs, it takes prece-
dence. The time interval depends on the time base.
Norm
The instrument acquires a waveform only if a trigger occurs. The
"Norm" LED lights up in green.
[Source]
Toggles the trigger source between the channels: C1...C8. The key lights up in the
color of the selected channel.
[Slope]
Toggles the trigger slope or trigger polarity, dependent on the trigger type. The current
setting is shown on the trigger label.
[Zone]
Opens a dialog to configure the zone trigger. A zone trigger combines the trigger condi-
tion with the intersection or non-intersection of one or more zones or masks.