Acquisition and waveform setup
R&S
®
MXO 5 Series
117User Manual 1802.3369.02 ─ 02
Remote command:
CHANnel<ch>:INVert on page 467
Label
Defines a label text. The label is shown at the waveform on the right edge of the dis-
play.
Skew
Sets a skew value to compensate for the delay of the measurement setup or from the
circuit specifics that the instrument cannot compensate automatically. It affects only the
selected input channel.
Delay differences between channels are caused by the different length of cables,
probes, and other sources. Correct deskew values are important for accurate trigger-
ing. Signals that are routed over lines with different lengths have a different propaga-
tion delay. This delay can lead to a non-synchronous waveform display.
Remote command:
CHANnel<ch>:SKEW:TIME on page 467
6.3.3 Bandwidth settings
Access: "Menu" > "Vertical" > "Bandwidth" tab
The "Bandwidth" tab provides all settings that affect the bandwidth of the measurement
system. The channels are listed in horizontal subtabs. Make sure to select the correct
channel tab before you enter the settings.
Bandwidth
Sets the bandwidth limit. The specified bandwidth indicates the range of frequencies
that the instrument can acquire and display accurately with less than 3 dB attenuation.
Frequencies above the limit are removed from the signal, and noise is reduced.
The selected bandwidth is shown on the signal icon.
For basic information, see also: "Bandwidth" on page 114.