Protocol analysis
R&S
®
MXO 5 Series
369User Manual 1802.3369.02 ─ 02
"Identifier"
Sets the trigger to one specific identifier or an identifier range. Enter
only the 6-bit identifier without parity bits, not the protected identifier.
Description of trigger type specific settings: Identifier setup: Condi-
tion, Frame ID min, Frame ID max.
"Identifier +
Data"
Sets the trigger to a combination of identifier and data condition. The
instrument triggers at the end of the last byte of the specified data
pattern.
The identifier conditions are the same as for the "Identifier" trigger
type, see Identifier setup: Condition, Frame ID min, Frame ID max.
Data conditions are set with Data setup: Condition, Pattern, Position.
"Wake-up
frame"
Triggers after a wake-up frame.
"Error condi-
tion"
Identifies various errors in the frame, see Error conditions.
Remote command:
TRIGger:LIN:TYPE on page 736
Identifier setup: Condition, Frame ID min, Frame ID max
The identifier setup consists of the condition and one or two identifier pattern.
Condition
Sets the comparison condition to a specific value or a range.
"Frame ID min"
Defines the bit pattern of the identifier. Enter only the 6-bit identifier
without parity bits, not the protected identifier.
"Frame ID
max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger:LIN:IMIN on page 738
TRIGger:LIN:IMAX on page 738